{"id":1725,"date":"2015-06-02T12:00:47","date_gmt":"2015-06-02T03:00:47","guid":{"rendered":"http:\/\/www.lookingatnothing.com\/?p=1725"},"modified":"2015-06-02T11:10:36","modified_gmt":"2015-06-02T02:10:36","slug":"investigations-the-case-of-the-incorrect-transmission-factors","status":"publish","type":"post","link":"https:\/\/lookingatnothing.com\/index.php\/archives\/1725","title":{"rendered":"Investigations: the case of the incorrect transmission factors"},"content":{"rendered":"<p>During the CanSAS meeting, a few comments pointed to the correct definition of transmission factor: &#8220;The ratio of the transmitted beam, <em>plus the scattered and diffracted radiation<\/em>, to the incident beam&#8221;.<\/p>\n<p>For SAXS experiments (but not Ultra-SAXS), we usually smile at that italicized part and wave it off as a small effect. But we may get ourselves into trouble with strongly scattering samples&#8230;<\/p>\n<p><!--more--><\/p>\n<p>Measuring a glassy carbon calibration sample should by now be a routine procedure in your toolkit. It is the one sample to keep close in times of metrological uncertainty. To measure glassy carbon correctly, however, one has to correct for the transmission: we need to compensate for the lack of detected radiation due to photon absorption in the sample. There are many ways of measuring the transmission factor of a sample <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=T_%7Bsample%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"T_{sample}\" class=\"latex\" \/>, with some ways better than others.<\/p>\n<p>According to the definition above, we can write:<\/p>\n<p><img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=T_%7Bsample%7D+%3D+%5Cfrac%7BI_1+%2B+I_s+%2B+I_d+%2B+I_i+%2B+%5Cldots%7D%7BI_0%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"T_{sample} = &#92;frac{I_1 + I_s + I_d + I_i + &#92;ldots}{I_0}\" class=\"latex\" \/><\/p>\n<p>With <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_1&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_1\" class=\"latex\" \/> the unaffected beam after the sample, <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_s&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_s\" class=\"latex\" \/> the scattered radiation, <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_d&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_d\" class=\"latex\" \/> diffracted radiation, <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_i&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_i\" class=\"latex\" \/> the inelastic scattering, etc. Basically every photon process <em>minus<\/em> absorption goes in the numerator.<\/p>\n<p>The definition above points to one good way of measuring transmission. If we are to place a large-aperture, high-dynamic range photon counter directly after the sample, we could measure both the photon flux of the direct beam as well as the intensity after the sample (<img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_1+%2B+I_s+%2B+&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_1 + I_s + \" class=\"latex\" \/> (some of) <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_d&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_d\" class=\"latex\" \/>). Such detectors, however, are either expensive or nonexistent (although some silicon drift detectors may surprisingly be capable of this).<\/p>\n<p>The next best thing would be a large area PIN-diode (I have some that are 10 <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=mm%5E2&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"mm^2\" class=\"latex\" \/>). These should give an output proportional to the intensities measured. However, this requires the installation of an additional detector, electronics and a mechanism to move the detector to directly downstream of the sample.<\/p>\n<figure id=\"attachment_1737\" aria-describedby=\"caption-attachment-1737\" style=\"width: 300px\" class=\"wp-caption alignright\"><a href=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_setup1.png\"><img loading=\"lazy\" decoding=\"async\" class=\"size-medium wp-image-1737\" src=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_setup1-300x87.png\" alt=\"Transmission measurement set-up in the Bruker Nanostar, in standard (case A) and modified (case B) form.\" width=\"300\" height=\"87\" srcset=\"https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_setup1-300x87.png 300w, https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_setup1-1024x296.png 1024w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/a><figcaption id=\"caption-attachment-1737\" class=\"wp-caption-text\"><strong>Figure 1:<\/strong> Transmission measurement set-up in the Bruker Nanostar, in standard (case A) and modified (case B) form.<\/figcaption><\/figure>\n<p>In our laboratory, however, we have a system that uses a sheet of 3mm thick glassy carbon (not calibrated) on an arm behind the sample (Figure 1). The idea then is that the scattering from that thick, strongly scattering sample is proportional to the intensity of the radiation impinging upon it. When the sample is not in place, the detected radiation is proportional to <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_0&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_0\" class=\"latex\" \/>, i.e.\u00a0<img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_%7Bdet%2C+0%7D+%3D+k+I_0+T_%7BGC%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_{det, 0} = k I_0 T_{GC}\" class=\"latex\" \/> (with T_{GC} the transmission factor of the 3mm glassy carbon). When the sample is in place, it <em>should<\/em> be proportional to <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_1&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_1\" class=\"latex\" \/> with the same factor <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=k&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"k\" class=\"latex\" \/>: <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_%7Bdet%2C+1%7D+%3D+k+I_0+T_%7BGC%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_{det, 1} = k I_0 T_{GC}\" class=\"latex\" \/>.<\/p>\n<p>One common problem I may have raised before is that if you have a strongly scattering sample, this may pass through the 3mm thick glassy carbon (or, alternatively, cause a second scattering event in there). That means that in that case <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_%7Bdet%2C+1%7D+%3D+T_%7BGC%7D+%28k+I_1+%2B+I_s+T_%7Bsample%7D%29+&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_{det, 1} = T_{GC} (k I_1 + I_s T_{sample}) \" class=\"latex\" \/>. Note that we do not know <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=k&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"k\" class=\"latex\" \/> (or <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=T_%7BGC%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"T_{GC}\" class=\"latex\" \/>), and so cannot figure out how much scattering comes from the sample and how much from the glassy carbon.<\/p>\n<figure id=\"attachment_1735\" aria-describedby=\"caption-attachment-1735\" style=\"width: 273px\" class=\"wp-caption alignright\"><a href=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing.png\"><img loading=\"lazy\" decoding=\"async\" class=\"size-medium wp-image-1735\" src=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing-273x300.png\" alt=\"Partial removal of the sample scattering from the 3mm glassy carbon pattern by means of a slit between sample and 3mm GC.\" width=\"273\" height=\"300\" srcset=\"https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing-273x300.png 273w, https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing-932x1024.png 932w, https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing.png 1586w\" sizes=\"auto, (max-width: 273px) 100vw, 273px\" \/><\/a><figcaption id=\"caption-attachment-1735\" class=\"wp-caption-text\"><strong>Figure 2:<\/strong> Partial removal of the sample scattering from the 3mm glassy carbon pattern by means of a slit between sample and 3mm GC.<\/figcaption><\/figure>\n<p>The easy solution for this is to install a collimator between the sample and glassy carbon, to prevent the sample scattering to impinge upon the glassy carbon or detector (Case B in Figure 1). This I have done long ago, and can be clearly seen in the transmission images obtained (Figure 2). We then use the counts recorded in the shadowed area only.<\/p>\n<p>So what is the problem then?<\/p>\n<p>The problem is our calibration sample. This calibration sample is a 1 mm piece of glassy carbon, whose scattering should be in accordance with a dataset measured on Jan Ilavsky&#8217;s instrument. Assuming a gravimetric density of 1.42 g\/cc and a composition consisting of C, such a piece <a href=\"http:\/\/11bm.xray.aps.anl.gov\/absorb\/absorb.php?mode=calc&amp;spectrumtype=Energy&amp;spectrum=17.4&amp;formula=C&amp;radius=0.5&amp;densitytype=RHO&amp;density=1.42&amp;submit=Compute\">should give us<\/a> a transmission factor of 0.955. Using our method, we get 0.90(1). Such a discrepancy I cannot accept.<\/p>\n<p>It could well be that we are missing intensity in our transmitted beam <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_1&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_1\" class=\"latex\" \/>. Such a loss in intensity could be due to scattering of a significant fraction of <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=I_0&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"I_0\" class=\"latex\" \/> by the sample. Whether or not this is the case, could be answered by calculating an ostensibly simple value: the fraction of the incident beam which is scattered by our sample.<\/p>\n<p>&#8230; and I thought I would have an answer for you by now. However, despite the simplicity of the posed question, I have yet to figure out how exactly this should be done. I am not even sure it <em>can<\/em> be done since we do not know the beam characteristics such as the degree of coherence. Hints and suggestions are welcome in the comments.<\/p>\n<p>&nbsp;<\/p>\n","protected":false},"excerpt":{"rendered":"<div class=\"mh-excerpt\"><p>During the CanSAS meeting, a few comments pointed to the correct definition of transmission factor: &#8220;The ratio of the transmitted beam, plus the scattered and <a class=\"mh-excerpt-more\" href=\"https:\/\/lookingatnothing.com\/index.php\/archives\/1725\" title=\"Investigations: the case of the incorrect transmission factors\">[&#8230;]<\/a><\/p>\n<\/div>","protected":false},"author":2,"featured_media":1735,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"activitypub_content_warning":"","activitypub_content_visibility":"","activitypub_max_image_attachments":4,"activitypub_interaction_policy_quote":"anyone","activitypub_status":"","footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2}},"categories":[1],"tags":[],"class_list":["post-1725","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-uncategorized"],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/lookingatnothing.com\/wp-content\/uploads\/2015\/06\/GC_shadowing.png","jetpack_shortlink":"https:\/\/wp.me\/p1gZ2v-rP","jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/1725","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/comments?post=1725"}],"version-history":[{"count":8,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/1725\/revisions"}],"predecessor-version":[{"id":1743,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/1725\/revisions\/1743"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/media\/1735"}],"wp:attachment":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/media?parent=1725"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/categories?post=1725"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/tags?post=1725"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}