{"id":2029,"date":"2016-02-23T10:45:03","date_gmt":"2016-02-23T09:45:03","guid":{"rendered":"http:\/\/www.lookingatnothing.com\/?p=2029"},"modified":"2016-02-24T16:32:11","modified_gmt":"2016-02-24T15:32:11","slug":"step-by-step-measuring-liquids-in-absolute-units","status":"publish","type":"post","link":"https:\/\/lookingatnothing.com\/index.php\/archives\/2029","title":{"rendered":"Step-by-step: measuring liquids in absolute units"},"content":{"rendered":"<p>[<em>apologies for my absence last week: I was felled by work and illness. The following is based on a request from a reader:<\/em>]<\/p>\n<p>Providing measurements in absolute units has several benefits. Amongst others, it allows you to determine the volume fraction of scatterers, and it lets you compare measurements better (thus helping to shield you from procedural mistakes). Furthermore, you score extra brownie points with reviewers of your papers, as it shows that some care has been taken in the collection. This has been very easy to do for solid samples, thanks to Jan Ilavsky&#8217;s glassy carbon samples. Now, though, we have liquid samples, so how do we do this in practice?<\/p>\n<p><!--more--><\/p>\n<h3>Prerequisites:<\/h3>\n<figure id=\"attachment_2035\" aria-describedby=\"caption-attachment-2035\" style=\"width: 300px\" class=\"wp-caption alignright\"><a href=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2016\/02\/6539302565_92eab42088_z.png\" rel=\"attachment wp-att-2035\"><img loading=\"lazy\" decoding=\"async\" class=\"size-medium wp-image-2035\" src=\"http:\/\/www.lookingatnothing.com\/wp-content\/uploads\/2016\/02\/6539302565_92eab42088_z-300x200.png\" alt=\"We deal in absolutes. Image: CC-licensed by Marina Aguiar https:\/\/flic.kr\/p\/aXRCPX\" width=\"300\" height=\"200\" srcset=\"https:\/\/lookingatnothing.com\/wp-content\/uploads\/2016\/02\/6539302565_92eab42088_z-300x200.png 300w, https:\/\/lookingatnothing.com\/wp-content\/uploads\/2016\/02\/6539302565_92eab42088_z.png 640w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/a><figcaption id=\"caption-attachment-2035\" class=\"wp-caption-text\">We deal in absolutes.<br \/>Image: CC-licensed by Marina Aguiar https:\/\/flic.kr\/p\/aXRCPX<\/figcaption><\/figure>\n<p>First off, what I would very strongly recommend for liquid samples is a flow-through capillary, or at the least a re-usable capillary. The standard (borosilicate, lindemann glass, etc.) <a href=\"http:\/\/www.hilgenberg-gmbh.de\/en\/catalog\/test-tubes\/mark-tubes\/mark-tubes-static-samples\/\">Mark-tubes<\/a> should not be used for this. While they are present in almost every lab as liquid sample containers, the variance between tubes is &#8212; in my experience &#8212; too large to be able to do a decent background subtraction. Only a few have succeeded in this, with many more leaving the field amidst blood-covered shards of these damn tubes. Furthermore, the inner diameter is not so homogeneous, so an accurate measure of the quantity of material in the beam is also lacking.<\/p>\n<p>We, fortunately, have in our lab a flow-through cell made from some magic plastic tubing that enters and exits our vacuum chamber. The plastic has a low enough scattering signal to not interfere with our measurements. Anton Paar sells this tube, though anyone with a Raman, IR or NMR instrument is also welcome to take a section and figure out what it is made of (as long as you report it here :)). However, any decent re-usable capillary or glass flow-through cell will do.<\/p>\n<h3>The procedure<\/h3>\n<p>Our current procedure to measure liquids in absolute units has three steps:<\/p>\n<p>First, We determine the transmission factors of everything we need: our capillary filled with air (or vacuum), a capillary filled with water, a capillary with our background solution and a capillary with our sample solution. Determining the transmission factor accurately, though, isn&#8217;t quite as easy as we thought (<a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1725\">1<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1904\">2<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1910\">3<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1949\">4<\/a>). The uncertainties in this determination are directly responsible for the uncertainty of your absolute scale.<\/p>\n<p>Second, We measure accurately the scattering pattern of our empty (or air-filled) capillary, followed by an accurate measurement of water. In a slit-focused system, 20 minutes for each is sufficient (since desmearing is not necessary for what we intend to do with this data). However, for pinhole-collimated systems, you may need a bit longer to get good data, perhaps a few hours or an overnight measurement.<\/p>\n<p>Both datasets are then corrected for the basics. That means <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1751\">DS, DC, MK, TI, FL, TR, SP,<\/a> and in that order (click for clarification on the abbreviations), where <a href=\"http:\/\/dx.doi.org\/10.1088\/0953-8984\/25\/38\/383201\">DC is the darkcurrent correction<\/a>. The difference of these two datasets will give you the scattering pattern of water. Water is supposed to have a largely flat scattering profile (see <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1910\">my previous post on this<\/a>), if it doesn&#8217;t, chances are you did something wrong in the data correction.<\/p>\n<p>At room temperature, this flat scattering level of water is supposed to lie at 1.62 <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=m%5E%7B-1%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"m^{-1}\" class=\"latex\" \/>. That scaling difference between your measured water level in your arbitrary units and 1.62 <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=m%5E%7B-1%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"m^{-1}\" class=\"latex\" \/> is exactly what you need. By scaling your future data with this value (the calibration factor), you can shift things to absolute units! Easy, right?<\/p>\n<p>So the third step is to measure your samples and sample background solutions. As long as you don&#8217;t change too much on your instrument (don&#8217;t realign the optics, change the capillary or let the dogs loose in the lab), the calibration factor shouldn&#8217;t change. So you measure your sample background solution, your sample itself, process both with <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1751\">DS, DC, MK, TI, FL, TR, SP<\/a>, subtract and multiply the resulting sample scattering pattern with your calibration factor. Done: absolute units.<\/p>\n<h3>In other words<\/h3>\n<p>So to reiterate, using one capillary, you:<\/p>\n<ol>\n<li>Measure what the transmission factors are of the empty capillary, the capillary with water, with your sample background solution and with your sample. If your instrument supports this, you may do this during the following measurements. Mind the pitfalls though (<a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1725\">1<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1904\">2<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1910\">3<\/a>, <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1949\">4<\/a>)!<\/li>\n<li>Measure the empty capillary, and that capillary filled with water.<\/li>\n<li>Apply the basic data corrections <a href=\"http:\/\/www.lookingatnothing.com\/index.php\/archives\/1751\">DS, DC, MK, TI, FL, TR, SP<\/a> (you can leave out SP for small angles, and FL for stable instruments).<\/li>\n<li>Subtract from one another to obtain a flat water scattering profile.<\/li>\n<li>Determine the calibration factor you need to use to bring the water level up to 1.62 <img decoding=\"async\" src=\"https:\/\/s0.wp.com\/latex.php?latex=m%5E%7B-1%7D&#038;bg=ffffff&#038;fg=000&#038;s=0&#038;c=20201002\" alt=\"m^{-1}\" class=\"latex\" \/><\/li>\n<li>Measure your sample background, and your sample<\/li>\n<li>Apply the basic data corrections (as point 3 above), and subtract the sample background from the sample measurement.<\/li>\n<li>Multiply the result with your calibration factor from point 5.<\/li>\n<li>Do a little victory dance: you rock! Doubly so if you propagated the uncertainties through all of this, in which case I also tip my hat to you.<\/li>\n<\/ol>\n<h3>What else?<\/h3>\n<p>There are optional additional steps to this. You can, if you want, include a flatfield (FF) and\/or thickness (TH) correction after background subtraction. The thickness correction may help a bit later on to keep the calibration factor largely independent of your capillary choice. If you need to, you can do a de-smearing step at the very end.<\/p>\n<p>All the corrections and abbreviations given here are detailed in my <a href=\"http:\/\/dx.doi.org\/10.1088\/0953-8984\/25\/38\/383201\">now ancient paper<\/a>.<\/p>\n","protected":false},"excerpt":{"rendered":"<div class=\"mh-excerpt\"><p>[apologies for my absence last week: I was felled by work and illness. The following is based on a request from a reader:] Providing measurements <a class=\"mh-excerpt-more\" href=\"https:\/\/lookingatnothing.com\/index.php\/archives\/2029\" title=\"Step-by-step: measuring liquids in absolute units\">[&#8230;]<\/a><\/p>\n<\/div>","protected":false},"author":2,"featured_media":2035,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"activitypub_content_warning":"","activitypub_content_visibility":"","activitypub_max_image_attachments":4,"activitypub_interaction_policy_quote":"anyone","activitypub_status":"","footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2}},"categories":[1],"tags":[],"class_list":["post-2029","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-uncategorized"],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/lookingatnothing.com\/wp-content\/uploads\/2016\/02\/6539302565_92eab42088_z.png","jetpack_shortlink":"https:\/\/wp.me\/p1gZ2v-wJ","jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/2029","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/comments?post=2029"}],"version-history":[{"count":5,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/2029\/revisions"}],"predecessor-version":[{"id":2038,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/posts\/2029\/revisions\/2038"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/media\/2035"}],"wp:attachment":[{"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/media?parent=2029"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/categories?post=2029"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/lookingatnothing.com\/index.php\/wp-json\/wp\/v2\/tags?post=2029"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}